Abstract
The aim of this study was to semi-automate process analytics for the quantification of common impurities in downstream processing such as host cell DNA, host cell proteins and endotoxins using a commercial liquid handling station. By semi-automation, the work load to fully analyze the elution peak of a purification run was reduced by at least 2.41 h. The relative standard deviation of results among different operators over a time span of up to 6 months was at the best reduced by half, e.g. from 13.7 to 7.1% in dsDNA analysis. Automation did not improve the reproducibility of results produced by one operator but released time for data evaluation and interpretation or planning of experiments. Overall, semi-automation of process analytics reduced operator-specific influence on test results. Such robust and reproducible analytics is fundamental to establish process analytical technology and get downstream processing ready for Quality by Design approaches.
Author supplied keywords
Cite
CITATION STYLE
Christler, A., Felföldi, E., Mosor, M., Sauer, D., Walch, N., Dürauer, A., & Jungbauer, A. (2020). Semi-automation of process analytics reduces operator effect. Bioprocess and Biosystems Engineering, 43(5), 753–764. https://doi.org/10.1007/s00449-019-02254-y