Multianalyte serum analysis using mid-infrared spectroscopy

160Citations
Citations of this article
53Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

This study assesses the potential for using mid-infrared (mid-IR) spectroscopy of dried serum films as the basis for the simultaneous quantitation of eight serum analytes: total protein, albumin, triglycerides, cholesterol, glucose, urea, creatinine and uric acid. Infrared transmission spectra were acquired for 300 serum samples, each analysed independently using accepted reference clinical chemical methods. Quantitation methods were based upon the infrared spectra and reference analyses for 200 specimens, and the models validated using the remaining 100 samples. Standard errors in the IR-predicted analyte levels (S(y/x)) were 2.8 g/L (total protein), 2.2 g/L (albumin), 0.23 mmol/L (triglycerides), 0.28 mmol/L (cholesterol), 0.41 mmol/L (glucose) and 1.1 mmol/L for urea, with correlation coefficients (IR vs reference analyses) of 0.95 or better. The IR method emerged to be less suited for creatinine (Sy/x = μmol/L) and uric acid (Sy/x = 140 μmol/L) due to the relatively low concentrations typical of these analytes.