A stability-indicating high performance liquid chromatographic assay for the simultaneous determination of atenolol and lercanidipine hydrochloride in tablets

10Citations
Citations of this article
15Readers
Mendeley users who have this article in their library.
Get full text

Abstract

A simple, rapid, precise and accurate isocratic reversed phase stability indicating HPLC method was developed and validated for the simultaneous determination of atenolol and lercanidipine hydrochloride in commercial tablets. The chromatographic separation was achieved on phenomenex Gemini C18 (250×4.6 mm, 5 μm) column using a mobile phase consisting of acetonitrile and buffer (20 mM potassium dihydrogen phosphate pH 3.5) in the ratio of (55:45, v/v) at a flow rate of 1.0 ml/min and UV detection at 235 nm. The linearity of the proposed method was investigated in the range of 40-160 μg/ml (r 2 =0.9995) for atenolol and 8-32 μg/ml (r 2 =0.9993) for lercanidipine. Degradation products produced as a result of stress studies did not interfere with the detection of atenolol and lercanidipine and the assay can thus be considered stability-indicating.

Cite

CITATION STYLE

APA

Kaila, H. O., Ambasana, M. A., Thakkar, R. S., Saravaia, H. T., & Shah, A. K. (2011). A stability-indicating high performance liquid chromatographic assay for the simultaneous determination of atenolol and lercanidipine hydrochloride in tablets. Indian Journal of Pharmaceutical Sciences, 73(4), 376–380. https://doi.org/10.4103/0250-474X.95612

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free