Abstract
Beam studies of thin single- and double-stage THGEM-based detectors are presented. Several 10 × 10 cm2 configurations with a total thickness of 5-6 mm (excluding readout electronics), with 1 × 1 cm 2 pads inductively coupled through a resistive layer to APV-SRS readout electronics, were investigated with muons and pions. Detection efficiencies in the 98% range were recorded with an average pad-multiplicity of ∼ 1.1. The resistive anode resulted in efficient discharge damping, with few-volt potential drops; discharge probabilities were ∼ 10-7 for muons and 10-6 for pions in the double-stage configuration, at rates of a few kHz/cm2. These results, together with the robustness of THGEM electrodes against spark damage and their suitability for economic production over large areas, make THGEM-based detectors highly competitive compared to the other technologies considered for the SiD-DHCAL. © CERN 2013.
Author supplied keywords
Cite
CITATION STYLE
Bressler, S., Arazi, L., Natal Da Luz, H., Azevedo, C. D. R., Moleri, L., Oliveri, E., … Breskin, A. (2013). Beam studies of novel THGEM-based potential sampling elements for Digital Hadron Calorimetry. Journal of Instrumentation, 8(7). https://doi.org/10.1088/1748-0221/8/07/P07017
Register to see more suggestions