Preliminary work of real-time ultrasound imaging system for 2-D array transducer

7Citations
Citations of this article
8Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Ultrasound (US) has emerged as a non-invasive imaging modality that can provide anatomical structure information in real time. To enable the experimental analysis of new 2-D array ultrasound beamforming methods, a pre-beamformed parallel raw data acquisition system was developed for 3-D data capture of 2D array transducer. The transducer interconnection adopted the row-column addressing (RCA) scheme, where the columns and rows were active in sequential for transmit and receive events, respectively. The DAQ system captured the raw data in parallel and the digitized data were fed through the field programmable gate array (FPGA) to implement the pre-beamforming. Finally, 3-D images were reconstructed through the devised platform in real-time.

Cite

CITATION STYLE

APA

Li, X., Yang, J., Ding, M., & Yuchi, M. (2015). Preliminary work of real-time ultrasound imaging system for 2-D array transducer. Bio-Medical Materials and Engineering, 26, S1579–S1585. https://doi.org/10.3233/BME-151457