How air influences radiation dose deposition in multiwell culture plates: A Monte Carlo simulation of radiation geometry

3Citations
Citations of this article
19Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Radiation of experimental culture cells on plates with various wells can cause a risk of underdosage as a result of the existence of multiple air-water interfaces. The objective of our study was to quantify this error in culture plates with multiple wells. Radiation conditions were simulated with the GAMOS code, based on the GEANT4 code, and this was compared with a simulation performed with PENELOPE and measured data. We observed a slight underdosage of ∼4% on the most superficial half of the culture medium. We believe that this underdosage does not have a significant effect on the dose received by culture cells deposited in a monolayer and adhered to the base of the wells.

Cite

CITATION STYLE

APA

Sabater, S., Berenguer, R., Honrubia-Gomez, P., Rivera, M., Nuñez, A., Jimenez-Jimenez, E., … Ramirez-Castillejo, C. (2014). How air influences radiation dose deposition in multiwell culture plates: A Monte Carlo simulation of radiation geometry. Journal of Radiation Research, 55(5), 1009–1014. https://doi.org/10.1093/jrr/rru022