Abstract
This study reports the results of a Japanese clinical trial of transcatheter closure of atrial septal defects (ASD) using the Amplatzer septal occluder (ASO). Thirty-five patients with secundum ASD underwent transcatheter closure using the ASO at a median age of 12,9 years (range, 3,2-29.2 years) and a median weight of 39.2kg (range, 11.6-65.1 kg). The ASO was successfully implanted in 34 patients. The mean ASD diameter of the 34 patients measured by transesophageal echocardiography was 11.7±4.2mm (range, 5.0-20.8mm) and the mean balloon stretched diameter was 16.8±4.2mm (range, 9-25mm). The mean ASO size was 16.9±4.3mm (range, 9-26mm). Complete closure rate at 1 day and 1 year after closure was 91% and 97%, respectively. One patient developed a transient second-degree atrioventricular block during the implantation procedure. No other complications occurred. Transcatheter closure of ASD using the ASO is effective and safe.
Author supplied keywords
Cite
CITATION STYLE
Oho, S., Ishizawa, A., Akagi, T., Dodo, Hi., & Kato, H. (2002). Transcatheter closure of atrial septal defects with the Amplatzer septal occluder: A Japanese clinical trial. Circulation Journal, 66(9), 791–794. https://doi.org/10.1253/circj.66.791
Register to see more suggestions