Abstract
The cellulose chromatography method of the Association of Official Analytical Chemists for determination of Aflatoxin M1 was compared with a two-dimensional thin layer chromatography procedure. The minimum detectable Aflatoxin M1 by the Official procedure was approximately .01 μg/kg while the thin layer chromatography procedure has a claimed sensitivity of .004 μg/kg. Analyses of milk samples by thin layer chromatography agreed with results of the Official procedure in the range of .01 to .25 μg/liter. Each method consistently yielded recoveries greater than 87%, and there was no significant difference between means or variances of each method. The thin layer chromatography method took less time than the Official method and showed a reproducible lower threshold sensitivity of .005 μg/liter. © 1977, American Dairy Science Association. All rights reserved.
Cite
CITATION STYLE
Bodine, A. B., Wayne Davis, R., Janzen, J. J., & Crawford, F. M. (1977). Comparison of Two Methods for Analysis of Aflatoxin M1 in Milk. Journal of Dairy Science, 60(3), 450–452. https://doi.org/10.3168/jds.S0022-0302(77)83886-1