Abstract
A rapid, simple technique combining disk solid-phase extraction and handheld X-ray fluorescence (XRF) spectrometry was developed for the on-site determination of As, Se, and Cr(VI) in drinking water. For the preconcentration of As, Se, and Cr(VI), a 50-mL aqueous sample was adjusted to pH 4, followed by passage through a Ti-loaded anion-exchange disk (Ti-AED). Both sides of the Ti-AED were coated with an adhesive cellophane tape prior to drying using a cordless hair iron, followed by a handheld XRF measurement. The Ti-AED adsorbed As(III), As(V), Se(IV), Se(VI), and Cr(VI) in water without requiring oxidation and reduction. The detection limits of As, Se, and Cr(VI) were all 1.0 μg L -1 . Good recoveries were obtained by measuring certified reference materials, and spiking natural mineral water with As, Se, and Cr(VI). As the proposed method does not require a power supply or toxic reagents in any analytical step, it is suitable for the on-site determination of toxic elements in drinking water.
Author supplied keywords
Cite
CITATION STYLE
Hagiwara, K., Koike, Y., Aizawa, M., & Nakamura, T. (2018). On-site determination of arsenic, selenium, and chromium(VI) in drinking water using a solid-phase extraction disk/handheld X-ray fluorescence spectrometer. Analytical Sciences, 34(11), 1309–1315. https://doi.org/10.2116/analsci.18P217