Radiophotoluminescence Imaging Reader for Passive Dosimetry

10Citations
Citations of this article
16Readers
Mendeley users who have this article in their library.

Abstract

In this study, we constructed an image reader system for two-dimensional radiation dosimetry based on radiophotoluminescence (RPL). The reader system mainly consists of a Peltier-cooled CMOS camera and an LED excitation source, as well as control units, and it offers an arbitrary excitation wavelength (e.g., 365, 405, 460, 530, and 630 nm) and a wide spectral detection range (200-1000 nm) to be used for a wide range of prototype image detectors having different spectral features. Using the developed system, we were able to successfully obtain X-ray projection images recorded on a commercial RPL detector [Ag-doped phosphate glass (Ag-PG) plate] as well as custom-made prototypes of flexible imaging plates (IPs). Furthermore, the reader system is applicable not only for RPL imaging but also for conventional optically stimulated luminescence (OSL) imaging, which was demonstrated by using a commercial IP for medical diagnosis.

Cite

CITATION STYLE

APA

Nanto, H., Okada, G., Hirasawa, K., Koguchi, Y., Shinozaki, W., Ueno, S., … Yamamoto, T. (2022). Radiophotoluminescence Imaging Reader for Passive Dosimetry. Sensors and Materials, 34(2), 757–764. https://doi.org/10.18494/SAM3686