Identification of predicate creep under the 510(k) process: A case study of a robotic surgical device

15Citations
Citations of this article
13Readers
Mendeley users who have this article in their library.
Get full text

Abstract

The FDA's 510(k) process for medical devices is based on “substantial equivalence” to devices clearedpre-1976 or legally marketed thereafter, known as predicate devices. In the last decade, several high-profile device recalls have drawn attention to this regulatory clearance process and researchers have raised questions about the validity of the 510(k) process as a broad clearance mechanism. One of the issues raised is the risk of predicate creep, a cycle of technology change through repeated clearance of devices based on predicates with slightly different technological characteristics, such as materials and power sources, or have indications for different anatomical sites. This paper proposes a new way to identify potential “predicate creep” through the use of product codes and regulatory classifications. We test this method by applying it to a case study of a Robotic Assisted Surgery (RAS) device, the Intuitive Surgical Da Vinci Si Surgical System. We find that there is evidence of predicate creep using our method, and discuss implications of this method for research and policy.

Cite

CITATION STYLE

APA

Lefkovich, C., & Rothenberg, S. (2023). Identification of predicate creep under the 510(k) process: A case study of a robotic surgical device. PLoS ONE, 18(3 March). https://doi.org/10.1371/journal.pone.0283442

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free