Comparison between rebound tonometer and Tono-Pen® in relation to Goldmann applanation tonometry and the influence of central corneal thickness on these three methods

2Citations
Citations of this article
5Readers
Mendeley users who have this article in their library.

Abstract

Objective: to compare intraocular pressure (IOP) measurements from two different instruments, the rebound tonometer (Icare®) and the Tono-Pen® XL (TP), with that from Goldmann applanation tonometer (GAT) and to determine the influence of central corneal thickness on these three instruments. Methods: 274 eyes from 137 healthy volunteers were evaluated in this cross-sectional study. All IOP measurements were made by the same examiner who was blinded to the observed result. Icare® tonometry was performed first, followed by TP, GAT, and pachymetry in a random order. Results: a good correlation was observed between IOP measurements obtained with Icare® and GAT (r = 0.79, p = 0.000) and between TP and GAT (r = 0.69, p = 0.000). Icare® and TP measurements were consistently higher than GAT measurements. A Bland-Altman plot indicated that the 95% limits of agreement between Icare® and GAT were 0.98 ± 3.12 (mean ± SD; range, -5.14 to 7.11) mmHg, and those between TP and GAT were 1.88 ± 3.20 (range, -4.38 to 8.15) mmHg. In the group of patients with thinner corneas, Icare® overestimated IOP by 0.5mmHg compared with IOP obtained using GAT and by 1.8 mmHg compared with IOP obtained using TP. In the group of patients with thicker corneas, Icare® overestimated IOP by 1.4 mmHg compared IOP obtained using GAT and by 1.5 mmHg compared with IOP obtained using TP. Conclusion: IOP measurements obtained with Icare® and TP showed a good correlation with that of GAT. Both tonometers tend to overestimate IOP compared to GAT. In patients with thinner corneas, Icare® perfrmed better than TP.

Cite

CITATION STYLE

APA

Magela-Vieira, G., Carvalho-Sousa, H. C., & Pinto-Silva, L. N. (2018). Comparison between rebound tonometer and Tono-Pen® in relation to Goldmann applanation tonometry and the influence of central corneal thickness on these three methods. Revista Mexicana de Oftalmologia, 92(6). https://doi.org/10.24875/RMOE.M18000034

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free