Abstract
To examine the biological effects of extremely low frequency magnetic field (ELFMF), we have designed and manufactured a new equipment for long-term and high-density exposure of cells to ELFMF. The ELFMF exposure system consists of a generator of magnets with a bult-in CO2incubator, an alternating current (AC) power supply, a gas compressor and a thermocontroller for the incubator, and a cooling unit for the magnets. The CO2incubator made of acrylic resin is inserted into the inner-space of the silicon steel strip-cores. In this system, the temperature of the incubator is maintained at 37±0.5°C. The maximum magnetic flux density on the exposure area of the incubator is 500 mT (T; tesla) at a current of 556 Arms (rms; root mean square) at 50 Hz. The long-term (up to 120 hr) exposure of 400 mT ELFMF did not affect the growth of both HL60RG and CCRF-CEM cells originated from human leukemia. The post-X-irradiation exposure of 400 mT ELFMF for 2 hr also did not affect the radiation sensitivity of GM0637 and TAT2SF cells originated from a normal human and an ataxia telangiectasia patient. © 1994, Journal of Radiation Research Editorial Committee. All rights reserved.
Cite
CITATION STYLE
Miyakoshi, J., Ohtsu, S., Tatsumi-MIYAJIMA, J., & Takebe, H. (1994). A Newly Designed Experimental System for Exposure of Mammalian Cells to Extremely Low Frequency Magnetic Fields. Journal of Radiation Research, 35(1), 26–34. https://doi.org/10.1269/jrr.35.26