A 31 T split-pair pulsed magnet for single crystal x-ray diffraction at low temperature

14Citations
Citations of this article
10Readers
Mendeley users who have this article in their library.
Get full text

Abstract

We have developed a pulsed magnet system with panoramic access for synchrotron x-ray diffraction in magnetic fields up to 31 T and at low temperature down to 1.5 K. The apparatus consists of a split-pair magnet, a liquid nitrogen bath to cool the pulsed coil, and a helium cryostat allowing sample temperatures from 1.5 up to 250 K. Using a 1.15 MJ mobile generator, magnetic field pulses of 60 ms length were generated in the magnet, with a rise time of 16.5 ms and a repetition rate of 2 pulses/h at 31 T. The setup was validated for single crystal diffraction on the ESRF beamline ID06. © 2014 AIP Publishing LLC.

Cite

CITATION STYLE

APA

Duc, F., Fabrèges, X., Roth, T., Detlefs, C., Frings, P., Nardone, M., … Rikken, G. L. J. A. (2014). A 31 T split-pair pulsed magnet for single crystal x-ray diffraction at low temperature. Review of Scientific Instruments, 85(5). https://doi.org/10.1063/1.4878915