Abstract
A flow injection system using a tubular electrode based on the redox properties of copper (II) ions occluded in EVA membrane was developed for L-ascorbic acid determination. The polyethylene-co-vinyl acetate (EVA 40% m/m) matrix was doped with copper (II) ions and dispersed on the surface of a graphite/epoxy tubular electrode. The electrode showed a super-Nernstian response for L-ascorbic acid concentration between 10-3 and 10-1 mol L-1 with a detection limit of 8.5×10-4 mol L-1, when 0.1 mol L-1 KH2PO4 buffer mixed with 0.1 mol L-1 hydrogen peroxide at pH 5.0 was employed as carrier. The Potentiometric-FIA system allows an analytical frequency of 120 samples per hour with a precision of 3.6%. The results obtained for ascorbic acid determination in pharmaceutical samples, without any previous treatment, were similar to those obtained by the British Pharmacopoeia method.
Author supplied keywords
Cite
CITATION STYLE
Fernandes, J. C. B., Rover, L., Kubota, L. T., & De Oliveira Neto, G. (2000). Potentiometric Determination of L-Ascorbic Acid in Pharmaceutical Samples by FIA Using a Modified Tubular Electrode. Journal of the Brazilian Chemical Society, 11(2), 182–186. https://doi.org/10.1590/S0103-50532000000200014