Pneumatic tube system induced haemolysis: Assessing sample type susceptibility to haemolysis

61Citations
Citations of this article
37Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Background: The pneumatic tube system (PTS) has been implicated in inducing haemolysis. It is not known whether certain sample types are more susceptible to haemolysis than others. We assessed the level of haemolysis in commonly used sample types in the clinical biochemistry department when transported through the PTS. Method: Blood was collected in pairs for different sample types and sent to the laboratory via the pneumatic tube or delivered by a porter. Haemolysis indices were measured spectrophotometrically and compared for each pair of sample type. Results: Our results suggest that plain serum samples are more susceptible to haemolysis than the other sample types when sent through our PTS (P <0.0001). Compared with serum with gel samples, plain serum samples are more prone to haemolysis (P<0.001). This suggests that gel may confer some protection against haemolysis. Conclusion: Different hospitals will have varying system configurations and use different sample types. We recommend that each hospital investigate their own system to assess whether haemolysis is a recurring problem in any of the sample types transported.

Cite

CITATION STYLE

APA

Sodi, R., Darn, S. M., & Stott, A. (2004). Pneumatic tube system induced haemolysis: Assessing sample type susceptibility to haemolysis. Annals of Clinical Biochemistry, 41(3), 237–240. https://doi.org/10.1258/000456304323019631

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free