Benchmarking the gamma pass score using ArcCHECK for routine dosimetric QA of VMAT plans

7Citations
Citations of this article
17Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

A minimum expected gamma (γ) pass rate for VMAT plan verification using ArcCHECK was established based on the RTTQA, TGI 19 test cases and 10 clinical plans with varying levels of complexity. The impact of the 'Measurement Uncertainty' parameter as available in the ArcCHECK software on γ pass rate was studied for both global and local γ analysis. Our results show that excluding measurement uncertainty adds tighter tolerance in local γ comparison. From the verification of our benchmark cases we established minimum expected γ pass rates of 85% and 88% for 2%/2mm global and 3%/3mm local tolerance criteria.

Cite

CITATION STYLE

APA

Arumugam, S., Young, T., Xing, A., Thwaites, D., & Holloway, L. (2015). Benchmarking the gamma pass score using ArcCHECK for routine dosimetric QA of VMAT plans. In Journal of Physics: Conference Series (Vol. 573). Institute of Physics Publishing. https://doi.org/10.1088/1742-6596/573/1/012040