Objective: To evaluate variables that affect risk of contamination for endoscopic retrograde cholangiopancreatography and endoscopic ultrasound endoscopes. Design: Observational, quality improvement study. Setting: University medical center with a gastrointestinal endoscopy service performing ∼1,000 endoscopic retrograde cholangiopancreatography and ∼1,000 endoscopic ultrasound endoscope procedures annually. Methods: Duodenoscope and linear echoendoscope sampling (from the elevator mechanism and instrument channel) was performed from June 2020 through September 2021. Operational changes during this period included standard reprocessing with high-level disinfection with ethylene oxide gas sterilization (HLD-ETO) was switched to double high-level disinfection (dHLD) (June 16, 2020-July 15, 2020), and duodenoscopes changed to disposable tip model (March 2021). The frequency of contamination for the co-primary outcomes were characterized by calculated risk ratios. Results: The overall pathogenic contamination rate was 4.72% (6 of 127). Compared to duodenoscopes, linear echoendoscopes had a contamination risk ratio of 3.64 (95% confidence interval [CI], 0.69-19.1). Reprocessing using HLD-ETO was associated with a contamination risk ratio of 0.29 (95% CI, 0.06-1.54). Linear echoendoscopes undergoing dHLD had the highest risk of contamination (2 of 18, 11.1%), and duodenoscopes undergoing HLD-ETO and the lowest risk of contamination (0 of 53, 0%). Duodenoscopes with a disposable tip had a 0% contamination rate (0 of 27). Conclusions: We did not detect a significant reduction in endoscope contamination using HLD-ETO versus dHLD reprocessing. Linear echoendoscopes have a risk of contamination similar to that of duodenoscopes. Disposable tips may reduce the risk of duodenoscope contamination.
Cite
CITATION STYLE
Ayres, A. M., Wozniak, J., O’Neil, J., Stewart, K., Leger, J. S., Pasculle, A. W., … Snyder, G. M. (2023). Endoscopic retrograde cholangiopancreatography and endoscopic ultrasound endoscope reprocessing: Variables impacting contamination risk. Infection Control and Hospital Epidemiology, 44(9), 1485–1489. https://doi.org/10.1017/ice.2022.319
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.