Automated pressure-controlled discography with constant injection speed and real-time pressure measurement

5Citations
Citations of this article
12Readers
Mendeley users who have this article in their library.

Abstract

Objective : This study was designed to investigate automated pressure-controlled discography (APCD) findings, to calculate the elastance of intervertebral discs, and to assess the relationship between the calculated elastance and disc degeneration. Methods : APCD was performed in 19 patients. There were a total of 49 intervertebral discs treated. Following intradiscal puncture, a dye was constantly injected and the intradiscal pressure was continuously measured. The elastance of the intervertebral disc was defined as unit change in intradiscal pressure per fractional change in injected dye volume. Disc degeneration was graded using a modified Dallas discogram scale. Results : The mean elastance was 43.0 ± 9.6 psi/mL in Grade 0, 39.5 ± 8.3 psi/mL in Grade 1, 30.5 ± 22.3 psi/mL in Grade 2, 30.5 ± 22.3 psi/mL in Grade 3, 13.2 ± 8.3 psi/mL in Grade 4 and 6.9 ± 3.8 psi/mL in Grade 5. The elastance showed significant negative correlation with the degree of degeneration (R2 = 0.529, p = 0.000). Conclusion : APCD liberates the examiner from the data acquisition process during discography. This will likely improve the quality of data and the reliability of discography. Elastance could be used as an indicator of disc degeneration. Copyright © 2009 The Korean Neurosurgical Society.

Cite

CITATION STYLE

APA

Kim, H. I., & Shin, D. A. (2009). Automated pressure-controlled discography with constant injection speed and real-time pressure measurement. Journal of Korean Neurosurgical Society, 46(1), 16–22. https://doi.org/10.3340/jkns.2009.46.1.16

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free