Phantom Study on Dose Reduction Using Iterative Reconstruction in Low-dose Computed Tomography for Lung Cancer Screening

3Citations
Citations of this article
17Readers
Mendeley users who have this article in their library.

Abstract

We investigated dose reduction ability of an iterative reconstruction technology for low-dose computed tomography (CT) for lung cancer screening. The Sinogram Affirmed Iterative Reconstruction (SAFIRE) provided in a multi slice CT system, Somatom Definition Flash (Siemens Healthcare) was used. An anthropomorphic chest phantom (N-1, Kyoto Kagaku) was scanned at volume CT dose index (CTDIvol) of 0.50-11.86 mGy with 120 kV. For noise (standard deviation) and contrast-to-noise ratio (CNR) measurements, CTP486 and CTP515 modules in the Catphan (The Phantom Laboratory) were scanned. Radiological technologists were participated in the perceptual comparison. SAFIRE reduced the SD values by approximately 50% compared with filter back projection (FBP). The estimated dose reduction rates by SAFIRE determined from the perceptual comparison was approximately 23%, while 75% dose reduction rate was expected from the SD value reduction of 50%.

Cite

CITATION STYLE

APA

Minehiro, K., Takata, T., Hayashi, H., Sakuda, K., Nunome, H., Kawashima, H., & Sanada, S. (2015). Phantom Study on Dose Reduction Using Iterative Reconstruction in Low-dose Computed Tomography for Lung Cancer Screening. Nihon Hōshasen Gijutsu Gakkai Zasshi, 71(12), 1201–1208. https://doi.org/10.6009/jjrt.2015_JSRT_71.12.1201