Abstract
How to improve placement accuracy of needle insertion into liver tissue is of paramount interest to physicians. A robot-assisted system was developed to experimentally demonstrate its advantages in needle insertion surgeries. Experiments of needle insertion into porcine liver tissue were performed with conic tip needle (diameter 8 mm) and bevel tip needle (diameter 1.5 mm) in this study. Manual operation was designed to compare the performance of the presented robot-assisted system. The real-time force curves show outstanding advantages of robot-assisted operation in improving the controllability and stability of needle insertion process by comparing manual operation. The statistics of maximum force and average force further demonstrates robot-assisted operation causes less oscillation. The difference of liver deformation created by manual operation and robot-assisted operation is very low, 1 mm for average deformation and 2 mm for maximum deformation. To conclude, the presented robot-assisted system can improve placement accuracy of needle by stably control insertion process.
Author supplied keywords
Cite
CITATION STYLE
Wang, W., Shi, Y., Goldenberg, A. A., Yuan, X., Zhang, P., He, L., & Zou, Y. (2015). Experimental analysis of robot-assisted needle insertion into porcine liver. Bio-Medical Materials and Engineering, 26, S375–S380. https://doi.org/10.3233/BME-151325