Abstract
Most Czerny-Turner based LIBS systems are used widely. In the present work, an instrumental setup based on Paschen-Runge is introduced, and analytical parameters are optimized for LIBS. With optimum parameters, analytical performance measures, such as precision, accuracy and detection limit, are compared for LIBS and Spark-OES. The results show that the analytical accuracy obtained by LIBS is similar to that obtained with Spark-OES, while the precision and detection limit are notably close to those of Spark-OES. Thus, LIBS can be used satisfactorily for process analysis in the metallurgy industry. © 2014 ISIJ.
Author supplied keywords
Cite
CITATION STYLE
Zhang, Y., Jia, Y., Chen, J., Shen, X., Liu, Y., Zhao, L., … Ma, H. (2014). Comparison of the analytical performances of laser-induced breakdown spectroscopy and spark-OES. ISIJ International, 54(1), 136–140. https://doi.org/10.2355/isijinternational.54.136