Reducing Glucose Meter Adverse Events by Using Reliability Growth With the FDA MAUDE Database

5Citations
Citations of this article
13Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Glucose meter evaluations are common and provide important information about glucose meter performance versus standards. Although some meters meet guidelines and others fall short in these evaluations, most results are within the A and B zones of a glucose meter error grid. Another data source that is seldom used is the FDA adverse event database (MAUDE). This database describes glucose meter malfunctions and injury as reported by actual users and returned 10 837 adverse events across all meters for the first 7 months of 2018. Reliability growth management is an established tool to reduce failure rates. A reliability growth example is presented followed by a discussion of how this tool could be applied to reduce glucose meter failure events using the MAUDE database.