An Immunosensor for the Determination of Cortisol in Serum and Saliva by Array SPRi

11Citations
Citations of this article
19Readers
Mendeley users who have this article in their library.

Abstract

Cortisol is a hormone which plays an essential role in the immune, endocrine, cardiovascular, renal and skeletal systems. Its level increases in response to stress, illness, injury or exhaustion, and it is therefore a significant diagnostic biomarker of stress. An immunosensor for the determination of cortisol by SPRi array was developed. The receptive part of the immunosensor is mouse monoclonal antibody against cortisol, immobilized via cysteamine linker. The optimum pH of the immunosensor is 7.4, and the optimum concentration of the antibody is 50 ng mL−1. The immunosensor is specific for cortisol, and its linear response ranges from 0.20 ng mL−1 (LOQ) to 8 ng mL−1. The precision of the determination was between 3.1% and 3.3%, and the recovery between 99% and 102%. The immunosensor was validated by simultaneous determination of cortisol in serum and saliva samples by a standard method, with good agreement between the results.

Author supplied keywords

Cite

CITATION STYLE

APA

Sankiewicz, A., Oldak, L., Zelazowska-Rutkowska, B., Hermanowicz, A., Lukaszewski, Z., & Gorodkiewicz, E. (2022). An Immunosensor for the Determination of Cortisol in Serum and Saliva by Array SPRi. Sensors, 22(24). https://doi.org/10.3390/s22249675

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free