Impact of infusion set characteristics on the accuracy of patient-controlled morphine administration: A controlled in-vitro study

7Citations
Citations of this article
14Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

The aim of our in-vitro study was to assess the impact of infusion set characteristics on the accuracy of morphine doses in patient-controlled analgesia. Two infusion sets differing in conception and dead-space volume were assessed: a standard set and a low dead-space volume Y-set. The patient-controlled analgesia programme parameters were as follows: bolus equal to 1 ml at 100 ml.h-1; lockout intervals equal to 5 and 10 min; and carrier fluid flow rate equal to 10 and 50 ml.h-1. Morphine concentration was determined by an ultraviolet spectrophotometric method. The morphine doses were significantly different from one set to the other during bolus and lockout intervals, whatever the patient-controlled analgesia programme. The average doses were approximately 1.3-6.0 times higher with the low dead-space volume Y-set during bolus. Our study underlines the impact of infusion set characteristics on the accuracy of morphine patient-controlled analgesia doses. © 2013 The Association of Anaesthetists of Great Britain and Ireland.

Cite

CITATION STYLE

APA

Foinard, A., Décaudin, B., Barthélémy, C., Lebuffe, G., Debaene, B., & Odou, P. (2014). Impact of infusion set characteristics on the accuracy of patient-controlled morphine administration: A controlled in-vitro study. Anaesthesia, 69(2), 131–136. https://doi.org/10.1111/anae.12523

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free