Abstract
Objective: The aim of this study is to develop a simple, precise and accurate analytical method of deoxyarbutin in anhydrous emulsion system preparation. Methods: The analysis was conducted using high-performance liquid chromatography (HPLC). Chromatographic analysis was carried out using a reversed phase-C18 column. The mobile consists of two phases methanol and water (60: 40 v/v) at a flow rate of 1.0 ml/min. The determinations were performed using UV detector set at 225 nm. All validation procedures were added with hydroquinone as an internal standard. Results: The method showed coefficient correlation is 0.9978, relative standard deviation (RSD) smaller than 2%, Limit of Detection (LOD) and Limit of Quantitation (LOQ) are 0.599 µg/ml and 1.817 µg/ml respectively. The total amount deoxyarbutin in anhydrous emulsion preparation is 1.964+0.02% with 98% recovery percentage. Conclusion: The developed HPLC analytical method meets the validation criteria made by International Conference on Harmonisation (ICH).
Author supplied keywords
Cite
CITATION STYLE
Muchtaridi, M., Musfiroh, I., & Fauzi, A. (2019). Method development and validation for analysis of deoxyarbutinin anhydrous emulsion system using high-performance liquid chromatography. International Journal of Applied Pharmaceutics, 11(2), 172–175. https://doi.org/10.22159/ijap.2019v11i2.30461
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.