Abstract
The Hitachi L-8500A is a newly available apparatus for amino acid (AA) analysis that allows automatic on-line mixing of the ninhydrin reagent. The within-run precision (human plasma pools at three different concentrations) showed CVs <3.8% except for the lowest concentration of citrulline (4.4%), Tyr (4.5%), and α-aminobutyric acid (7.6%), and for the intermediate concentration of Asp (8.7%). Between-run precision (CV) was <3.1% for 17 AAs and <8.0% for 24 of 25 AAs (CV Asp = 12.0%). For retention times, within-run precision was <0.4% and between-run precision <1.8%. Excellent relations were found between the results from the Hitachi L-8500A and the widely used Beckman 6300 analyzer (0.929≤ r≤0.999). The detection was still linear at 5 μmol/L except for Pro and hydroxyproline (20 μmol/L). The upper limit was at least 2500 μmol/L for 13 AAs and at least 1000 μmol/L for 27 of 29 AAs (anserine = 500, Val = 600 μmol/L). Values from 100 human plasma samples agreed with previously published data. We conclude that the results obtained with the Hitachi L-8500A are satisfactory when compared with those of other AA analyzers utilizing the same method. Furthermore, the Hitachi L-8500A displays several advantages including programming flexibility, microsample capacity, low noise plotting, ammonia filtering, and manual repacking of the analytical column.
Author supplied keywords
Cite
CITATION STYLE
Le Boucher, J., Charret, C., Coudray-Lucas, C., Giboudeau, J., & Cynober, L. (1997). Amino acid determination in biological fluids by automated ion-exchange chromatography: Performance of Hitachi L-8500A. Clinical Chemistry, 43(8), 1421–1428. https://doi.org/10.1093/clinchem/43.8.1421
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.