Standardising the quantification of trifluoroacetic acid in water: interlaboratory validation trial using liquid chromatography-mass spectrometric detection (LC–MS/MS)

3Citations
Citations of this article
14Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

The concentration of trifluoroacetic acid in surface and groundwaters and drinking water is subject of increasing concern. Since there is no standardised procedure available, German standardisation body DIN has initiated the first draft standard using liquid chromatography with tandem mass spectrometry (LC–MS/MS) after direct injection for the concentration range between ≥ 0.1 and 3 µg/L. Herein, the interlaboratory comparison as final validation step involving 12 expert laboratories from Germany and Switzerland is described. Two groundwaters, two surface waters, a drinking water, a rainwater, and two fortified groundwater samples displayed relative repeatability standard deviations between 3 and 7% and relative reproducibility standard deviation between 6 and 20% and were included in the draft standard DIN 38407-53.

Cite

CITATION STYLE

APA

Dorgerloh, U., Becker, R., Freeling, F., Scheurer, M., Gökçe, E., Strack, K., … Ruth, K. (2025). Standardising the quantification of trifluoroacetic acid in water: interlaboratory validation trial using liquid chromatography-mass spectrometric detection (LC–MS/MS). Accreditation and Quality Assurance, 30(4), 473–476. https://doi.org/10.1007/s00769-025-01640-2