Gamma-ray efficiency of a HPGe detector as a function of energy and geometry

20Citations
Citations of this article
51Readers
Mendeley users who have this article in their library.
Get full text

Abstract

The concept of double points detector model approach (DPDM) is developed as a procedure to find the full energy peak efficiency of the coaxial 120cm3 closed hyperpure germanium (HPGe) detectors. Usually in the experimental nuclear physics work, which involves using HPGe detector for gamma-ray spectrometry, the full energy peak efficiency function must represent adequately the HPGe detector response. In the current work the gamma-ray energy in the range from 60keV to 1332keV and gamma-ray intensity changes by changing source to detector distance from 10mm to 800mm. The detector was characterized using a number of point-like standard sources. The calculated efficiencies obtained by (DPDM) are in good agreement with experimental data. © 2013 Elsevier Ltd.

Cite

CITATION STYLE

APA

Challan, M. B. (2013). Gamma-ray efficiency of a HPGe detector as a function of energy and geometry. Applied Radiation and Isotopes, 82, 166–169. https://doi.org/10.1016/j.apradiso.2013.07.010

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free