Meeting the New FDA Standard for Accuracy of Self-Monitoring Blood Glucose Test Systems Intended for Home Use by Lay Users

15Citations
Citations of this article
31Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

The OneTouch Verio Reflect blood glucose monitor (BGM) has market clearance in several countries based in part on fulfilling the lay user and system accuracy criteria described in ISO15197:2015. However, the Food and Drug Administration (FDA) does not recognize the accuracy criteria in ISO15197 as a basis for gaining regulatory clearance for these devices. The current study evaluates the BGM using the accuracy guidelines issued by the agency for self-monitoring blood glucose test systems for over-the-counter use. Glucose results were accurate vs comparator over a wide glucose range and met lay user and glucose accuracy criteria at extreme glucose values as described in the FDA guidance. Clinicaltrials.gov NCT03851549.

Cite

CITATION STYLE

APA

Katz, L. B., Stewart, L., King, D., & Cameron, H. (2020). Meeting the New FDA Standard for Accuracy of Self-Monitoring Blood Glucose Test Systems Intended for Home Use by Lay Users. Journal of Diabetes Science and Technology, 14(5), 912–916. https://doi.org/10.1177/1932296820906184