Possibility of large volume injection and band focusing in UHPLC

4Citations
Citations of this article
18Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

New shell-type stationary phases are widely used in fast chromatographic measurements. These columns provide more efficient separation, when applied in a conventional high-performance liquid chromatography instrument, than columns with fully porous particles, and the volume overload of core-shell particles is 60% of the value obtained for fully porous particles. Additionally, to achieve adequate sensitivity, the injection volume cannot be significantly decreased.This study presents a systematic evaluation of the possibilities of large volume injection onto columns packed with 2.6 μm Kinetex C18 shell particles. The effect of volume overload on performance of columns with different lengths (50, 100 and 150 mm) is studied. Column efficiency is compared under isocratic, pulse gradient and gradient conditions. The application of large volume injection in practice is also reported. The most suitable among the tested large volume injection techniques was the gradient elution, which was applied to determine amino acid enantiomers from fruit juice. © 2012 © The Author [2012]. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

Cite

CITATION STYLE

APA

Oláh, E., Tarnai, M., & Fekete, J. (2013). Possibility of large volume injection and band focusing in UHPLC. Journal of Chromatographic Science, 51(9), 839–844. https://doi.org/10.1093/chromsci/bms179

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free