Abstract
Atorvastatin is the primary choice for dyslipidemia treatment. Due to patent expiration of atorvastatin, the pharmaceutical industry makes copy of the drug. Therefore, the development methods for tablet quality tests involving atorvastatin concentration on tablets needs to be performed. The purpose of this research was to develop and validate the simple atorvastatin tablet analytical method by HPLC. HPLC system used in this experiment consisted of column Cosmosil C18 (150 x 4,6 mm, 5 μm) as the stationary reverse phase chomatography, a mixture of methanol-water at pH 3 (80:20 v/v) as the mobile phase, flow rate of 1 mL/min, and UV detector at wavelength of 245 nm. Validation methods were including: selectivity, linearity, accuracy, precision, limit of detection (LOD), and limit of quantitation (LOQ). The results of this study indicate that the developed method had good validation including selectivity, linearity, accuracy, precision, LOD, and LOQ for analysis of atorvastatin tablet content. LOD and LOQ were 0.2 and 0.7 ng/mL, and the linearity range were 20 - 120 ng/mL.
Cite
CITATION STYLE
Yugatama, A., Rohmani, S., & Dewangga, A. (2018). Development and Validation of High Performance Liquid Chromatography Method for Determination Atorvastatin in Tablet. In IOP Conference Series: Materials Science and Engineering (Vol. 333). Institute of Physics Publishing. https://doi.org/10.1088/1757-899X/333/1/012081