Abstract
Elevation of intracranial pressure is one of the most important issues in neurosurgery and neurology in clinical practice. The prevalent techniques for measuring intracranial pressure require equipments that are wired, restricted to a hospital environment, and cause patient discomfort. A novel method for measuring the intracranial pressure is described. A wireless completely implantable device, operating at an industrial-scientific-medical band of 2.4 GHz, has been developed and tested. In-vitro and in-vivo evaluations are described to demonstrate the feasibility of microwave pressure monitoring through scalp, device integrity over a long period of time, and repeatability of pressure measurements. A distinction between an epidural and sub-dural pressure monitoring techniques is also described. Histo-pathological results obtained upon a long-term device implantation favor the utilization of the sub-dural pressure monitoring method. On the other hand, in-vivo studies illustrate a maximum pressure reading error of 0.8 mm · Hg obtained for a sub-dural device with a capacitive microelectromechanical system sensor compared to 2 mm · Hg obtained for an epidural device with a piezoresistive sensor. © 2008 IEEE.