Investigation of sCVD diamond detectors for low energy heavy-ion reactions

3Citations
Citations of this article
1Readers
Mendeley users who have this article in their library.
Get full text

Abstract

The performance characteristics of single-crystalline diamond detectors at varying rates of heavy-ions below 50 MeV/u have been investigated. When tested with a 228Th source, the energy resolution of these detectors is better than 1.0% FWHM for 8.8 MeV α-particles. With a slight reduction in the energy resolution, simultaneous timing resolution of 314 ps has been achieved. Isotopic resolution has been achieved up to magnesium using a diamond telescope at low rates with a 20 MeV/u 20Ne + 12C reaction. At moderate rates of direct 7.5 MeV/u 78Kr, polarization effects are observed, though this effect can be diminished by increasing the applied voltage on the detector up to a limit. After an accumulated dose of 6.9 × 108 direct 78Kr particles transmitting through the detector, a degradation in the energy resolution of the 8.8 MeV peak from 1.6% to 5.3% is observed, indicating that permanent damage has been done to the detector.

Cite

CITATION STYLE

APA

Abbott, A., Sorensen, M., Tobin, Z., McIntosh, A. B., Hannaman, A., Gauthier, J., … Yennello, S. J. (2022). Investigation of sCVD diamond detectors for low energy heavy-ion reactions. Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1034. https://doi.org/10.1016/j.nima.2022.166796