Robust design of pulse oximeter using dynamic control and motion artifact detection algorithms

2Citations
Citations of this article
13Readers
Mendeley users who have this article in their library.

Abstract

Arterial oxygen saturation (SpO2) monitoring for newborns requires special attention in neonatal intensive care units (NICUs). Newborns have very low photo-plethysmogram (PPG) amplitudes and their body movements are difficult to contain. Hardware design and its associated signal processing algorithms should be robust enough so that faulty measurements can be avoided. In this study, improved designs were implemented to deal with low perfusion, motion artifact, and the influence of ambient light. Dynamic range was increased by using different LED intensities and a feedback system. To minimize the effects of motion artifact and to discard other unqualified data, four additional algorithms were used, which were based on dual-trace detection, continuity of DC level, morphology of PPG, and simultaneity check of SpO2. Our SpO2 system was tested with newborns with normal respiration in the NICU. Our system provided fast, real-time responses and 100% artifact detection was accomplished under 84% of SpO2.

Cite

CITATION STYLE

APA

Cho, J. H., Kim, J. C., & Yoon, G. W. (2014). Robust design of pulse oximeter using dynamic control and motion artifact detection algorithms. Journal of Electrical Engineering and Technology, 9(5), 1780–1787. https://doi.org/10.5370/JEET.2014.9.5.1780