Pulse contour analysis: Is it able to reliably detect changes in cardiac output in the haemodynamically unstable patient?

21Citations
Citations of this article
35Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Three pulse contour systems for monitoring cardiac output - LiDCO Plus™, PiCCO Plus™ and FloTrac™ - were compared in postcardiac surgery patients. None of the three methods demonstrated good trending ability according to concordance analysis. Pulse contour systems remain unreliable in the haemodynamically unstable patient. © 2011 BioMed Central Ltd.

Cite

CITATION STYLE

APA

Critchley, L. A. H. (2010, January 19). Pulse contour analysis: Is it able to reliably detect changes in cardiac output in the haemodynamically unstable patient? Critical Care. BioMed Central Ltd. https://doi.org/10.1186/cc9381