The performances of seven Immulite 2000 (Siemens Healthcare Diagnostics) TORCH (Toxoplasma gondii, other microorganisms, rubella virus, cytomegalovirus, and herpes simplex virus) assays were evaluated in comparison with the performances of the ETI-MAX 3000 (DiaSorin) TORCH assays. The two systems demonstrated good agreement, and given their sensitivity, specificity, and positive predictive value, they can be used with confidence for TORCH prenatal screening. Copyright © 2013, American Society for Microbiology. All Rights Reserved.

Cite

CITATION STYLE

APA

Centonze, A. R., Tonolli, E., & Fontana, R. (2013). Performance characteristics of current-generation immulite 2000 TORCH assays. Clinical and Vaccine Immunology, 20(1), 122–126. https://doi.org/10.1128/CVI.00106-12