Using integrated absorption to calibrate optical cavity spectrometers

5Citations
Citations of this article
11Readers
Mendeley users who have this article in their library.

Abstract

Sensitive absorption techniques using optical cavities (such as CEAS or ICOS) generally need the spectrometer response to be calibrated for quantitative measurements. Most calibrations are based on the instrument response to a known, steady state absorption. Such calibrations often have drawbacks in terms of cost, complexity, or convenience, especially for field measurements. In this paper, we show that the relationship between the integrated absorption and a known amount of absorber provides an alternative calibration strategy that yields a highly linear calibration curve and has a low uncertainty. This method is straightforward to implement and offers a practical alternative to other calibration strategies. This journal is

Cite

CITATION STYLE

APA

Fullam, D. P., Shoji, K., & Venables, D. S. (2015). Using integrated absorption to calibrate optical cavity spectrometers. Analytical Methods, 7(7), 3298–3301. https://doi.org/10.1039/c5ay00080g