Abstract
Study design: Circulating tumor cells is important in the clinical diagnosis of cancer and there are a number of circulating tumor cell detection systems associated with different isolation strategies being validated. There is a novel platform, the CytoBot 2000, which utilizes a combination of physical and immunological technologies to isolate and capture circulating tumor cells. Methods: In this retrospective study, 39 lung cancer patients and 11 normal healthy individuals were enrolled and performed circulating tumor cell tests and immunofluorescence staining with CytoBot 2000. The performance of this device was assessed by receiver operating characteristic curve. The clinical relevance of circulating tumor cells was assessed by Chi-square. The correlations between circulating tumor cell number and blood lymphocytes and tumor biomarkers were analyzed by Pearson correlation coefficient. Results: The number of circulating tumor cell is significantly increased in lung cancer patients (3.74 > 0.45, P
Author supplied keywords
Cite
CITATION STYLE
Mei, Z. F., Yan, J., Qian, L., He, Y. C., Feng, J. J., Wang, J., … Zou, H. (2023). Enrichment of Circulating Tumor Cells of Lung Cancer and Correlation With Serum Leukomonocyte and Tumor Biomarkers: A Retrospective Study. Technology in Cancer Research and Treatment, 22. https://doi.org/10.1177/15330338231167827