A self-focusing field emission (FE) X-ray tube with a large-area cathode design was simulated and fabricated. The designed X-ray tube had a cylindrically symmetric geometry; the diameter of the cathode and the anode was 15 mm, and the cathode-anode distance was 20 mm. Owing to the unique cup-shaped design of the cathode, the electron beam emitted from the large-area cathode was focused onto the anode without using magnetic lenses or extra biased electrodes. Carbon nanocoils, which were grown on the bottom of the circular cup-shaped cathode, were used as electron emitters because of their excellent FE properties. A simulation of the electron trajectories for various cup heights revealed that the optimal focal spot size (0.1 mm) was obtained at a cup height of 5 mm when a voltage of 50 kV was applied. To verify this result, an X-ray tube was fabricated and tested. An X-ray photograph of the tested tooth and electric circuits showed good resolution and X-ray intensity. The large cathode area effectively reduces the current density and thereby improves the lifetime of the cathode.
CITATION STYLE
Cheng, K., Chou, Y. P., Youh, M. J., Liu, Y. M., Ger, M. D., Hou, K. H., & Pu, N. W. (2015). Fabrication and simulation of self-focusing field emission X-ray tubes. Applied Sciences (Switzerland), 5(4), 942–954. https://doi.org/10.3390/app5040942