Abstract
This work investigates the behavior of commercial and custom Quartz tuning forkss (QTF) under humidity variations. The QTFs were placed inside a humidity chamber and the parameters were studied with a setup to record the resonance frequency and quality factor by resonance tracking. The variations of these parameters that led to a 1% theoretical error on the Quartz Enhanced Photoacoustic Spectroscopy (QEPAS) signal were defined. At a controlled level of humidity, the commercial and custom QTFs present similar results. Therefore, commercial QTFs appear to be a very good candidates for QEPAS as they are also affordable and small. When the humidity increases from 30 to 90 %RH, the variations in the custom QTFs’ parameters remain suitable, while commercial QTFs show unpredictable behavior.
Author supplied keywords
Cite
CITATION STYLE
Ayache, D., Rousseau, R., Kniazeva, E., Charensol, J., Seoudi, T., Bahriz, M., … Vicet, A. (2023). Commercial and Custom Quartz Tuning Forks for Quartz Enhanced Photoacoustic Spectroscopy: Stability under Humidity Variation. Sensors, 23(6). https://doi.org/10.3390/s23063135