Effects of E-BEAM sterilization on drug-eluting stents: Paclitaxel degradation elucidated by LC-MS-MS with information-dependent acquisition

3Citations
Citations of this article
7Readers
Mendeley users who have this article in their library.

Abstract

Effects of sterilization by electron beam (E-BEAM) on paclitaxel (1) mixed with poly(DL-lactide-co-glycolide) (PLG) in reservoirs of COSTAR Stents are examined by using liquid chromatography-mass spectrometry (LC-MS-MS) techniques with information-dependent acquisition (IDA). Numerous degradation products of 1 are formed in a β-radiation dose-dependent manner to give plethora of lowlevel degradants. This behavior, together with multiple interferences from PLG-related compounds, creates considerable challenges for analysis of the drug/PLG mixtures. IDA methods with different survey scans are proven to be very efficient in elucidating degradation pathways and in identifying numerous products. Combined LC-MS-MS results from analysis of sterilized drug substance and stents indicate that water addition and oxidative processes together with the isomerization are largely responsible for degradation of 1 under E-BEAM sterilization conditions used.

Cite

CITATION STYLE

APA

IlIchev, Y. V., & Alquier, L. (2011). Effects of E-BEAM sterilization on drug-eluting stents: Paclitaxel degradation elucidated by LC-MS-MS with information-dependent acquisition. Journal of Chromatographic Science, 49(10), 807–817. https://doi.org/10.1093/chrsci/49.10.807