Abstract
The recently developed PTR-TOF instrument was evaluated to measure methanol fluxes emitted from grass land using the eddy covariance method. The high time resolution of the PTR-TOF allowed storing full mass spectra up to 315 with a frequency of 10 Hz. Three isobaric ions were found at a nominal mass of 33 due to the high mass resolving power of the PTR-TOF. Only one of the three peaks contributed to eddy covariance fluxes. The exact mass of this peak agrees well with the exact mass of protonated methanol 33.0335). The eddy covariance methanol fluxes measured with PTR-TOF were compared to virtual disjunct eddy covariance methanol fluxes simultaneously measured with a conventional PTR-MS. The methanol fluxes from both instruments show excellent agreement.
Cite
CITATION STYLE
Müller, M., Graus, M., Ruuskanen, T. M., Schnitzhofer, R., Bamberger, I., Kaser, L., … Hansel, A. (2010). First eddy covariance flux measurements by PTR-TOF. Atmospheric Measurement Techniques, 3(2), 387–395. https://doi.org/10.5194/amt-3-387-2010