Reliability of isometric knee extension muscle strength measurements made by a hand-held dynamometer and a belt: A comparison of two types of device

22Citations
Citations of this article
67Readers
Mendeley users who have this article in their library.

Abstract

[Purpose] The purpose of the present study was to compare the reliability of 2 hand-held dynamometers (HHD-1, 2) with different designs, by performing isometric knee muscle extension measurements two times each. [Subjects] The subjects were 40 young healthy adults. [Methods] The reliability of the measurements was examined using Bland-Altman analysis. [Results] Bland-Altman analysis found a fixed bias in measurements made by HHD-1 with an average limits of agreement (LOA) value of −2.1 kgf. For HHD-2, only random errors were detected, and the minimal detectable change (MDC) was 11.4 kgf. Fixed biases were observed between the two devices with an average LOA value of 2.2 kgf. When the bodyweight ratio was used, fixed biases were observed in measurements made by both devices, and the average value of LOA was −0.03 kgf/kg. The comparison of the two devices revealed only random errors, and MDC was 0.22 kgf/kg. [Conclusion] For HHD measurements using these two devices, the appropriate number of measurements is two times, and comparison of measurement values between the two devices should be avoided.

Cite

CITATION STYLE

APA

Katoh, M. (2015). Reliability of isometric knee extension muscle strength measurements made by a hand-held dynamometer and a belt: A comparison of two types of device. Journal of Physical Therapy Science, 27(3), 851–854. https://doi.org/10.1589/jpts.27.851

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free