The Accuracy of Continuous Glucose Monitoring in the Medical Intensive Care Unit

25Citations
Citations of this article
23Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Objective: To assess the accuracy of continuous glucose monitoring (CGM) in medical intensive care unit (MICU) patients. Methods: A Medtronic Enlite® sensor accuracy was assessed versus capillary blood glucose (CBG) and plasma glucose (PG) using the mean absolute relative difference (MARD), surveillance error grid (SEG) analysis and modified Bland-Altman plots. Results: Using CBG as a reference, MARD was 6.6%. Overall, 99.7% of the CGM readings were within the “no risk” zone. No significant differences in accuracy were seen within vasopressor subgroups. Using PG as the reference, MARD was 8.8%. The surveillance error grid analysis showed 95.2% of glucose readings were within the “no risk” zone. There were no device-related adverse events. Conclusion: The CGM sensor showed acceptable accuracy in MICU patients, regardless of vasopressor use.

Cite

CITATION STYLE

APA

Tingsarat, W., Buranasupkajorn, P., Khovidhunkit, W., Boonchaya-anant, P., & Laichuthai, N. (2022). The Accuracy of Continuous Glucose Monitoring in the Medical Intensive Care Unit. Journal of Diabetes Science and Technology, 16(6), 1550–1554. https://doi.org/10.1177/19322968211027590