Abstract
Background: No recent clinical data on the incidence of catheter-related adverse events under insulin pump therapy have been reported. Methods: This was a prospective, two-period, observational, multicenter study in 45 diabetes outpatients (mean continuous subcutaneous insulin infusion [CSII] use, 6 years; mean hemoglobin A1c, 7.7%, at baseline). During the initial 1-month period (P1), the patients used their current catheter model, including a soft cannula in 98% of cases. They moved then to the new Accu-Chek® FlexLink catheter model (Disetronic Medical Systems AG, Burgdorf, Switzerland) for a 3-month period. The primary end point, including insertion failures and unexplained hyperglycemia within the first 6 h after catheter placement, was assessed from logbook records during P1 and the last month of the second period (P2). Secondary end points were catheter replacements for unexplained hyperglycemia and/or events at risk for immediate insulin delivery failure after the first 6 h. Results: Forty-five initial infusion failures occurred in 14 patients among 507 catheter insertions (8.9% of cases) during P1, whereas 15 similar events were seen in nine patients during P2 among 488 catheter insertions (3.1% of cases) (P
Cite
CITATION STYLE
Renard, E., Guerci, B., Leguerrier, A. M., & Boizel, R. (2010). Lower rate of initial failures and reduced occurrence of adverse events with a new catheter model for continuous subcutaneous insulin infusion: Prospective, two-period, observational, multicenter study. Diabetes Technology and Therapeutics, 12(10), 769–773. https://doi.org/10.1089/dia.2010.0073