Electrochemical concentration cell ozonesconde accuracy and precision.

71Citations
Citations of this article
5Readers
Mendeley users who have this article in their library.
Get full text

Abstract

The accuracy and precision of electrochemical concentration cell ozonesondes using 1.5% KI sensing solution were measured as a function of pressure altitude from 800 to 6 mbar. Measurements were made in an environmental chamber designed to monitor the instruments' response while subjecting them to ozone, pressure, and temperature profiles typical of those encountered during balloon-borne soundings. Data from dual-instrument balloon soundings were used to provide an independent estimate of precision. The precision was found to be 6-10% (one standard deviation) from 800 to 200 mbar, 5-6% from 200 to 10 mbar, and thereafter increasing to 16% at 6 mbar. The accuracy profile indicated a 3-5% positive error from 300 to 50 mbar. Larger positive errors were observed from 800 to 300 mbar (8-14%) and from 50 to about 15 mbar (10%). The error curve then rapidly rolled off toward negative values, reaching - 17% at 6 mbar. Use of new pumping efficiency measurement techniques should improve both accuracy and precision by 2-3% at the lowest pressures but would increase the accuracy error in the 50- to 15-mbar range by about the same amount.-Author

Cite

CITATION STYLE

APA

Barnes, R. A., Bandy, A. R., & Torres, A. L. (1985). Electrochemical concentration cell ozonesconde accuracy and precision. Journal of Geophysical Research, 90(D5), 7881–7887. https://doi.org/10.1029/JD090iD05p07881

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free