Determination of rare earth elements in ferrocarbonatite using ICP-AES and ICP-MS

5Citations
Citations of this article
16Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

A method has been developed for the determination of Rare Earth Elements (REEs) in ferrocarbonatite using Inductively Coupled Plasma—Atomic Emission Spectrometer (ICP-AES) and Inductively Coupled Plasma—Mass Spectrometer (ICP-MS). The conventional dissolution procedures are tedious, with large measurement uncertainty and are thus not suitable for certified reference material (CRM) production and exploration of REEs. The described method involves leaching of sample using 3 M HCl followed by dissolution of residue (silica and REEs) by HF then determined REEs using ICP-AES and ICP-MS. The leaching step prevents formation of fluoride precipitates of matrix (Ca, Mg, Ba and Al). The method has been validated using CRMs.

Cite

CITATION STYLE

APA

Thangavel, S., Venkateswarlu, G., Durgaprasad, A., Sunilkumar, B., Khuntia, A., Dash, K., & Durani, S. (2025). Determination of rare earth elements in ferrocarbonatite using ICP-AES and ICP-MS. Journal of Radioanalytical and Nuclear Chemistry, 334(6), 4251–4255. https://doi.org/10.1007/s10967-024-09771-5