Intermittent optical frequency measurements to reduce the dead time uncertainty of frequency link

48Citations
Citations of this article
20Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

The absolute frequency of the 87Sr lattice clock transition was evaluated with an uncertainty of 1.1 × 10-15 using a frequency link to the international atomic time (TAI). The frequency uncertainty of a hydrogen maser used as a transfer oscillator was reduced by homogeneously distributed intermittent measurement over a five-day grid of TAI. Three sets of four or five days measurements as well as systematic uncertainty of the clock at 8.6 × 10-17 have resulted in an absolute frequency of 87Sr 1S0-3P0 clock transition to be 429 228 004 229 872.85 (47) Hz.

Cite

CITATION STYLE

APA

Hachisu, H., & Ido, T. (2015). Intermittent optical frequency measurements to reduce the dead time uncertainty of frequency link. Japanese Journal of Applied Physics, 54(11). https://doi.org/10.7567/JJAP.54.112401