The packaging and clean method contribute to insulation failure of electrosurgical instruments

2Citations
Citations of this article
6Readers
Mendeley users who have this article in their library.
Get full text

Abstract

With the rapid development of medical technology, the use of electrosurgical instruments is dramatically increased in various types of surgery. However, the damage of the insulation layer of the reusable electrosurgical instrument often causes surgical accidents. The procedures of packaging and cleaning contribute to many damages to insulating layer of reusable electrosurgical instruments.Various types of reusable electrosurgical instruments were detected for insulation failures, conduction failures, short-circuit by using a high-voltage detector, DIATEG (Morgate company). In addition, reusable electrosurgical instruments were detected for insulation failures after packaging and cleaning by different procedures.13.1% (129/740) electrosurgical instruments had an insulation test failure; 6.2% (9/146) monopolar wires were with conduction failure; and 7.7% (16/207) bipolar wires were with short-circuit. Different packaging and cleaning procedures contribute to various degrees of damages to insulating property of reusable electrosurgical instruments.Insulation failure was a wide problem of reusable electrosurgical instruments, while fixed packaging method and mild cleaning procedures result in fewer damages to insulating property of reusable electrosurgical instruments.

Cite

CITATION STYLE

APA

Zhang, Y., Zhang, Y., Wang, Y., Yang, L., Hu, R., & Shalaby, M. N. (2021). The packaging and clean method contribute to insulation failure of electrosurgical instruments. Medicine (United States), 100(42), E27492. https://doi.org/10.1097/MD.0000000000027492

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free